Remote Control Commands
R&S
®
SMBV100A
732
Operating Manual 1176.8016.02 ─ 17
Example:
FM:EXT:COUP AC
selects the coupling mode AC for the external frequency modu-
lation signal.
Manual operation:
See
[:SOURce<hw>]:FM:EXTernal:DEViation
<Deviation>
Sets the deviation of the external frequency modulation signal in Hz. The maximum
deviation depends on the set RF frequency and the selected modulation mode (see
data sheet).
The sum of the deviations of all active frequency modulation signals may not exceed
the total value set with command
Parameters:
<Deviation>
float
Range:
see data sheet
Increment: 0.01
*RST:
1000
Example:
FM:EXT:DEV 3kHz
sets 3 kHz deviation to the frequency modulation signal.
Manual operation:
See
[:SOURce<hw>]:FM:INTernal:DEViation
<Deviation>
Sets the deviation of the internal frequency modulation signal in Hz.
The sum of the deviations of all active frequency modulation signals may not exceed
the total value set with command
Parameters:
<Deviation>
float
Range:
0 to dynamic
Increment: 0.01
*RST:
1E3
Example:
FM:INT1:DEV 2E3
sets 2 kHz deviation for the freqeuncy modulation signal.
Manual operation:
See
[:SOURce<hw>]:FM:MODE
<Mode>
Selects the mode for the frequency modulation.
SOURce Subsystem
Содержание SMBV100A
Страница 1: ...R S SMBV100A Vector Signal Generator Operating Manual Operating Manual 1176 8016 02 17 ÚÞ 2 ...
Страница 31: ...Contents R S SMBV100A 16 Operating Manual 1176 8016 02 17 ...
Страница 129: ...Manual Operation R S SMBV100A 114 Operating Manual 1176 8016 02 17 Legend of Front Panel Controls ...