Remote Control Commands
R&S
®
SMBV100A
700
Operating Manual 1176.8016.02 ─ 17
Return values:
<Clock>
float
Range:
0 to Max
Increment: 1E-3
*RST:
0
Example:
BB:MCCW:CLOC?
queries the output clock rate.
Response:
256 000 000
the output clock rate is 256
MHz.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:EXECute
Adopts the settings for the carrier range which has been defined using
the
:BB:MCCW:EDIT:CARR:...
commands.
Example:
BB:MCCW:EDIT:CARR:STAR 4
the carrier range starts at carrier 4.
BB:MCCW:EDIT:CARR:STOP 400
the carrier range stops at carrier 400.
BB:MCCW:EDIT:CARR:STAT ON
sets all the carriers in the carrier range to ON.
BB:MCCW:EDIT:CARR:EXEC
adopts the settings for all the carriers in the carrier range.
BB:MCCW:STAT
starts generation of the multi carrier signal. Carriers 4 to 400 are
in the ON state.
Usage:
Event
Manual operation:
See
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:PHASe:STEP
<Step>
Sets the step width by which the start phases of the carriers in the defined carrier
range will be incremented.
The phase settings are only valid if optimization of the crest factor is disabled
(
:SOURce:BB:MCCW:CFACtor:MODE OFF
).
Parameters:
<Step>
float
Range:
-359.99 to 359.99
Increment: 0.01
*RST:
0
SOURce Subsystem