5. Enter a password with a length of at least five digits.
This password is used on the remote computer to access the instrument. Other
settings may be changed according to the user-specific security requirements.
After the installation the Ultr@VNC program is automatically started together with
the operating system. On mouse over, the IP address of the instrument is indica-
ted.
This IP address and the user-defined password are the prerequisites to enable
remote access on the remote computer. Terminated connection is indicated by
changed icon color.
Installing the VNC Viewer on a Windows PC
1. Download the Ultr@VNC program form internet and follow the installation instruc-
tions.
Only the program component VNC Viewer is required.
Note:
The VNC Viewer program is included in the download for the installation of
the Ultr@VNC program on the signal generator if "Full installation" was selected in
the "Select Component" panel. In this case, the program
can be copied to the Windows PC.
2. Install VNC Viewer program component on the remote computer.
Setting up the VNC connection on the Linux/Unix remote computer
The VNC program is available per default for Linux/Unix operating systems.
1. Start a Web browser on the remote computer and enter the IP address of the
instrument.
2. Enter the following address:
vnc://<IP-address of the instrument>, e.g.
vnc://192.168.1.1
.
A dialog is opened and the password for the remote VNC connection is requested.
Remote Access via an External Controller