Remote Control Commands
R&S
®
SMBV100A
627
Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:DELay[:STARt]
<Start>
The command sets the start delay for the individual carriers in the defined carrier
range. If the command
:BB:ARB:MCAR:EDIT:CARR:DEL:STEP
is used to define a
step width, the delay entered here applies only to the starting carrier. The delays of the
remaining carriers are stepped up or down by the delay value specified in
the
:BB:ARB:MCAR:EDIT:CARR:DEL:STEP
command.
Parameters:
<Start>
float
Range:
0 to 1
Increment: 1E-9
*RST:
0
Default unit: s
Example:
BB:ARB:MCAR:EDIT:CARR:DEL 5us
sets a start delay of 5 us for the carriers in the carrier range.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:EXECute
The command adopts the settings for the carrier range which has been defined using
the
:BB:ARB:MCAR:EDIT:CARR:...
commands.
Example:
BB:ARB:MCAR:EDIT:CARR:STAR 4
the carrier range starts at carrier 2.
BB:ARB:MCAR:EDIT:CARR:STOP 20
the carrier range stops at carrier 20.
BB:ARB:MCAR:EDIT:CARR:STAT ON
sets all the carriers in the carrier range (2 to 20) to ON.
BB:ARB:MCAR:EDIT:CARR:EXEC
transfers the assistant settings for carrier 2 to 20 into the carrier
table.
Usage:
Event
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:FILE
<File>
Selects input file
Parameters:
<File>
string
Example:
BB:ARB:MCAR:EDIT:CARR:FILE "/var/user/IQ_wcdma'
Manual operation:
See
SOURce Subsystem