Instrument Function
R&S
®
SMBV100A
335
Operating Manual 1176.8016.02 ─ 17
Figure 5-23: Impact of the ramp settings
1
= "Ramp Function"
2, 3 = "Rise Delay", "Fall Delay"
4
= "Ramp Time"
5.9.2.6
Trigger Signals
In the R&S
SMBV, trigger signals are internally generated or externally supplied sig-
nals which start signal generation at a particular point in time.
Signal generation can also take place without triggering, in which case the signal is
then generated in full after modulation is powered up. A trigger event either has no
effect on signal generation (menu setting "Trigger Mode Auto") or triggers a signal
restart (menu setting "Trigger Mode Retrigger").
If signal generation is triggered, the signal is continuously generated after the first trig-
ger. In the "Armed_Auto" mode, a further trigger event has no effect. In the
"Armed_Retrig" mode, every additional trigger event triggers a restart of the signal. In
both cases, triggering can be reset to the initial state ("armed"), i.e. signal generation is
stopped and the instrument waits for the next trigger to start signal generation anew.
The status of signal generation ("Running" or "Stopped") is displayed for all trigger
modes in the corresponding trigger menu of the digital standard currently switched on.
The signal generation status particularly with an external trigger can thus be checked.
Internal and external trigger sources are available for triggering.
●
Internally, triggering is carried out manually by pressing the "Execute Trigger" but-
ton (menu selection "Internal").
●
External trigger signals can be fed in via the TRIGGER connector (menu selection
"External").
In the case of external triggering and internal triggering by the trigger signal on the
second path, the start can be delayed by a definable number of symbols (menu
setting "Trigger Delay External"). This makes it possible to purposely simulate the
time delay between a base station signal and a user equipment signal, for example
in the course of base station tests.
The effect of a restarted trigger signal in the "Retrigger" trigger mode can be sup-
pressed for a definable number of symbols (menu setting "External Trigger
Inhibit"). By this means the trigger can be suppressed for a definable number of
Baseband Signal - Baseband Block