Remote Control Commands
R&S
®
SMBV100A
716
Operating Manual 1407.6062.32 ─ 08
The instrument can also be remote controlled via the command set of several other gen-
erators, e.g. HP generator.
Please check for the corresponding Application Note at the download area of the product
site on the Internet. The Application Note is also provided on the user documentation CD-
ROM (included in delivery).
Note
: While working in a emulation mode, the instrument's specific command set is dis-
abled, i.e. the SCPI command
SYST:LANG
will be discarded.
The return to the SCPI command set of the R&S SMBV can only be performed by using
the appropriate command of the selected command set. For example, the HP command
EX
returns to the instrument-specific GPIB command set (selection
SYST:LANG
HPxxxx
).
Setting parameters:
<Language>
string
Example:
SYST:LANG?
Response: SCPI
The SCPI command set is used.
Usage:
Setting only
:SYSTem:ORESponse
<Oresponse>
The command defines the OPT string for selection user defined (
SYST:IDEN USER
).
Note
: While working in a emulation mode, the instrument's specific command set is dis-
abled, i.e. the SCPI command
SYST:ORES
will be discarded.
Parameters:
<Oresponse>
string
Example:
SYST:IDEN USER
selects an user-defined identification
SYST:ORES "Test Option"
defines the OPT string 'test option'
*OPT?
Response: 'test option'
:SYSTem:PROTect<ch>[:STATe]
<State>,<password>
The command activates and deactivates the specified protection level.
Suffix:
<ch>
.
Indicates the protection level.
There are several protection levels which disable specific service
functions (authorized personnel of R&S Service Departments
only).
Parameters:
<State>
0 | 1 | OFF | ON
SYSTem Subsystem
Содержание SMBV100A
Страница 31: ......
Страница 117: ...Manual Operation R S SMBV100A 100 Operating Manual 1407 6062 32 08 Legend of Front Panel Controls ...