Remote Control Commands
R&S
®
SMBV100A
524
Operating Manual 1407.6062.32 ─ 08
Example:
BB:ARB:MCAR:CARR:CONF?
queries the multi carrier conflict state.
Response:
0
no conflict has occurred.
Usage:
Query only
[:SOURce<hw>]:BB:ARB:MCARrier:CARRier<ch>:DELay
<Delay>
The command sets the start delay of the selected carrier.
Parameters:
<Delay>
float
Range:
0 to 1
Increment: 1E-9
*RST:
0
Default unit: s
Example:
BB:ARB:MCAR:CARR15:DEL 5us
sets a start delay of 50 us for carrier 15.
[:SOURce<hw>]:BB:ARB:MCARrier:CARRier<ch>:FILE
<File>
The command selects the file with I/Q data to be modulated onto the selected carrier.
Parameters:
<File>
<file name>
Example:
BB:ARB:MCAR:CARR15:FILE '/var/smbv/IQ_wcdma'
selects file
IQ_wcdma
. The data of the file is modulated onto car-
rier 15.
[:SOURce<hw>]:BB:ARB:MCARrier:CARRier<ch>:FREQuency
<Frequency>
Sets or indicates the carrier frequency, depending on the selected carrier frequency
mode.
The carrier frequnecy can be set in "Arbitrary Carrier frequency" mode. For "Equidistant
Carrier Spacing", the carrier spacing is determined automatically.
Parameters:
<Frequency>
integer
Range:
-40.0E6 to 40.0E6
Increment: 1
*RST:
0
Example:
BB:ARB:MCAR:CARR:MODE ARB
activates ARB multi carrier mode.
BB:ARB:MCAR:CARR:FREQ 5.0
sets 5.0 MHz carrier frequency.
SOURce Subsystem