Instrument Function
R&S
®
SMBV100A
389
Operating Manual 1407.6062.32 ─ 08
PEP
Enters the peak level of the external analog or digital baseband signal relative to full scale
of 0.5 V (in terms of dB full scale).
Indicates the peak level acquired with "Auto Level Set".
SCPI command:
Level
Indicates the estimated rms level acquired with "Auto Level Set".
SCPI command:
Signal Monitoring
The section "Signal Monitoring" indicates the overload of the A/D
-
Converter.
Overload
Indicates that the I/Q input is currently overloaded. This indication also appears in the
block diagram close to the I/Q connector of the Baseband Input block.
If overload is indicated either the amplitude of the external signal is too high (full scale of
0.5 V) and must be reduced or the entered "Peak Level" (in dB full scale) value does not
correspond with the real value and must be corrected. It also can be evaluated automat-
ically with button "Auto Level Set".
SCPI command:
Overload Hold
Indicates an overload since last reset for evaluating the measurement.
The Overload Hold state can be reset via the "Reset" button next to the LED, or is reset
automatically by starting a new measurement ("Auto
Level Set") or by setting new values
(Crest Factor, PEP, Level).
SCPI command:
Reset
Executes a reset of the Overload Hold State and LED.
SCPI command:
Connected Device
Indicates the ID and the transmitter sample rate, if supported by the transmitter.
External Baseband Signal - Baseband Input