Instrument Function
R&S
®
SMBV100A
388
Operating Manual 1407.6062.32 ─ 08
"OFL = Over-
flow"
The external sample rate frequency is too high which causes an over-
flow in signal transmission.
"URUN =
Underrun"
The external sample rate frequency is too low which causes an under-
run in signal transmission.
SCPI command:
Baseband Input Level
The crest factor and the peak power of the external baseband signal are entered in the
section "Baseband Input Level". These values are necessary for a correct internal signal
processing in the R&S Signal Generator. They can also be automatically estimated by a
measurement using the "Auto Level Set" button.
DIG IQ Auto Setting
Activates automatic adjustment of the baseband input signal.
The R&S SMBV receives PEP, level and crest factor values directly from the connected
transmitter and recognizes changes automatically.
SCPI command:
Measurement Period
Sets the recording duration for measuring the baseband input signal by
"Auto
Level
Set". For accurate level measurements, the measurement period should be
set to a time value that is long enough to capture several periods of the input signal.
SCPI command:
Auto Level Set
Starts measuring the input signal. The measurement estimates the crest factor, the peak
power and rms power. The estimated values are automatically entered into the input fields
"Crest
Factor", "PEP" and "Level". Using these estimated values the internal gain control
adjusts the input signal gain to achieve an optimal dynamic range.
SCPI command:
Crest Factor
Sets the crest factor of the external analog or digital baseband signal. Indicates the crest
factor acquired with "Auto Level Set".
SCPI command:
External Baseband Signal - Baseband Input