Instrument Function
R&S
®
SMBV100A
259
Operating Manual 1407.6062.32 ─ 08
AWGN Settings
The RFI signal generator is activated, the mode selected and the interfering signal is
configured in the top section of the menu.
State - AWGN
Activates/deactivates the RFI signal generator. The interferer (AWGN or CW interferer,
depending on the selected mode) is generated as after the generator is activated.
SCPI command:
Mode - AWGN
Selects the mode for generating the interfering signal.
"Additive
Noise"
The AWGN noise signal with selectable system bandwidth is added to
the baseband signal.
"Noise Only"
The pure AWGN noise signal with selectable system bandwidth is
modulated to the carrier. The connection to the baseband is interrupted.
"CW Interferer" A sine with a defined frequency offset is added to the baseband signal.
The calculation of E
b
/N
0
ratio is omitted.
SCPI command:
System Bandwidth - AWGN
(Additive Noise and Noise Only)
Sets the RF bandwidth to which the set carrier/noise ratio relates.
Within this frequency range, the signal is superimposed with a noise signal whose level
corresponds exactly to the set C/N or S/N ratio.
Note:
The noise signal is not generated arbitrarily for a particular bandwidth, but instead
it is generated in steps. Noise therefore also occurs outside the set system bandwidth.
SCPI command:
Minimum Noise/System Bandwidth Ratio - AWGN
(Additive Noise and Noise Only)
Sets the ratio of minimum noise bandwidth to system bandwidth.
Note:
The noise signal is not generated arbitrarily for a particular bandwidth, but instead
it is generated in steps. Noise therefore also occurs outside the set system bandwidth.
With this parameter the minimum real noise bandwidth can be set as required by some
standards. It does not influence the calculation of level from the set C/N or S/N ratio in
relation to system bandwidth.
The overall bandwidth is calculated as follow and may not exceed 120 MHz.
Overall Bandwidth = System BW x Minimum Noise/System BW Ratio
SCPI command:
Impairment of Digital I/Q Signal and Noise Generator - AWGN/IMP Block