Instrument Function
R&S
®
SMBV100A
106
Operating Manual 1407.6062.32 ─ 08
Risk of invalid adjustment
In order to achieve correct adjustment of the instrument, make sure that the instrument
is warm before performing adjustments. The warm-up time is 30 minutes.
To access the "Internal Adjustments" dialog, press the SETUP or MENU key under
"System".
The adjustments offered also depend on the installed options.
Adjust All
Starts all internal adjustments for which no external measuring equipment is needed. The
adjustments with external measuring equipment are described in the Service Manual
(supplied).
SCPI command:
Adjust Synthesis
Performs all adjustments which affect the frequency.
SCPI command:
Adjust Level
Performs all adjustments which affect the level. The acquired correction values improve
the settling time and the signal quality.
SCPI command:
Adjust I/Q Modulator Current Frequency
Starts the adjustment for the I/Q modulator for the currently set frequency, I/Q swap, and
baseband gain. The I/Q modulator is adjusted with respect to carrier leakage, I/Q imbal-
ance and quadrature.
General Instrument Settings