Operating Manual, 11/2007
6-153
E-1147.2759.00
Entering the DC Offset Voltage Component
Use
All generator functions can be combined with a DC voltage component.
Selecting
the channel
1.
Press the numeric keys
1
,
2
, or
3
to select channel
Ch 1
,
Ch 2
, or both
channels
Ch 1&2
.
The selected channel is displayed in green in the channel display.
Entering the
DC offset
2.
Press the
function key
in the
menu.
An entry field containing the currently applicable setting is displayed. The
default setting is “100 mV (0.1 FS)”. At the same time, the function keys
[13] are assigned various units of measurement.
Analog
Digital
3.
Enter a new value (
Ê
The permissible entry range is:
-10.6 V
≤
DC OFFSET
≤
10.6 V
(Analog)
-0.9999 FS
≤
DC OFFSET
≤
0.9999 FS
(Digital)
The new setting is stored and displayed in the parameter field.
(Analog)
(Digital)
Note:
The DC Offset function is only effective if combined with a signal
function (e.g. SINE). In this case, the maximum signal amplitude of the DC
OFFSET function depends on the amplitude of the other generator function;
the maximum signal amplitude decreases accordingly.
Ch 1
Ch 2