Upconverter Settings
R&S
®
SGU100A
86
User Manual 1176.7726.02 ─ 09
7.5
External Local Oscillator Settings
In a setup where an R&S SGU is connected to an incompatible signal generator (local
oscillator), the controller has to change the frequency and the level of this local oscilla-
tor (LO) to output the desired signal from the R&S
SGU.The required settings of the
local oscillator are shown in the "Local Oscillator" dialog.
► To access this dialog select "SGMA-GUI > Instrument Name >Ext. Local Oscilla-
tor".
When changing any parameters which affect the output signal of the combined system
of an incompatible LO and an R&S
SGU, follow these steps:
Setting the signal of the R&S SGU
1. Set the parameters of the R&S
SGU like frequency and level in the R&S SGMA-
GUI as needed.
2. Open the "SGMA-GUI > Instrument Name >Ext. Local Oscillator" dialog and read
the settings for "Frequency" and "Level".
3. Apply the required changes to the LO.
4. Press the "Confirm External LO Settings" button to confirm that you have made the
settings on the LO.
The desired signal is output by the R&S
SGU.
Frequency
Shows the desired frequency for the LO input signal.
Remote command:
[:SOURce]:LOSCillator:FREQuency?
Level
Shows the desired level for the LO input signal.
Remote command:
Confirm External LO Settings
Outputs the signal.
External Local Oscillator Settings