Remote Control Commands
R&S
®
SGU100A
175
User Manual 1176.7726.02 ─ 09
Usage:
Event
Manual operation:
See
[:SOURce]:POWer:ATTenuation:RFOFf:MODE
<Mode>
Selects the state which the attenuator assumes when the instrument is switched on.
Parameters:
<Mode>
MAX | FIXed
MAX
Sets attenuation to maximum when the RF signal is switched off.
This setting is recommended for applications that require a high
level of noise suppression.
FIXed
Retains the current setting and keeps the output impedance
constant during RF off.
*RST:
MAX
Manual operation:
See
[:SOURce]:POWer:ATTenuation:SOVer[:OFFSet]
<Offset>
Sets the switch-over offset value of the attenuator.
Parameters:
<Offset>
float
*RST:
0
Manual operation:
See
[:SOURce]:POWer:LMODe
<LevMode>
Selects the level mode.
Parameters:
<LevMode>
NORMal | LNOise | LDIStortion
NORMal
automatic selection of the best settings
LNOISe
settings for lowest noise
LDIStortion
settings for lowest distortions
*RST:
n.a. (factory preset: NORMal)
Example:
Chapter 11.1.3, "Advanced Task for Optimizing Perform-
Manual operation:
See
SOURce:POWer Subsystem