Remote Control Commands
R&S
®
Scope Rider RTH
375
User Manual 1326.1578.02 ─ 12
<n>
1..64
Order of harmonic (possibly restricted, see
Return values:
<State>
ON | OFF
ON
A limit violation occurred - the limit was exceeded.
OFF
No limit violation occurred.
Example:
HARM:LIM:TYPE EN50160
HARM:RES1:HARM2:LIMC?
//Result: 1
//The EN50160 defines a limit of 2% for the
//second order harmonic.
HARM:RES1:HARM2:LIM?
//Result: 2
HARM:RES1:HARM2:LIMV?
//Result: 1
//The calculated value exceeds the defined
// limit for the second harmonic.
HARM:RES1:LIMV?
//Result: 1
//Since the limit check for the second harmonic
// failed, the limit check for the entire
// signal failed.
Usage:
Query only
HARMonic:RESult<m>:HARMonics<n>:PHASe?
Queries the calculated phase value for the specified harmonic.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
<n>
1..64
Order of harmonic (possibly restricted, see
Return values:
<Phase>
Default unit: degrees
Usage:
Query only
HARMonic:RESult<m>:HARMonics<n>:RELMagnitude:MAXimum?
HARMonic:RESult<m>:HARMonics<n>:RELMagnitude:MINimum?
HARMonic:RESult<m>:HARMonics<n>:RELMagnitude[:CURRent]?
Returns the currently measured, minimum or maximum power level of the selected
harmonic, relative to the power level of the fundamental frequency.
Spectrum Analysis