Display Harmonics, Max No Harmonics
Determines the number of harmonics to be displayed in the bargraph. Up to 64 har-
monics can be displayed. By decreasing the number of harmonics, the bargraph
becomes easier to analyze.
Note that this setting only affects the graphical result display. It has no effect on the
measurement results or the selection of harmonics for which results are provided dur-
ing a file export (manually or via remote command). For that purpose, see
Harmonics that were previously eliminated by the
setting cannot be displayed.
To restrict the number of harmonics to display, select a predefined selection, or "User"
and enter the "Max No Harmonics". All harmonics of an order smaller than or equal to
the specified value (and selected for analysis) are displayed.
Remote command:
HARMonic:RESult<m>:NOHarmonics?
Average
Determines the number of acquisitions for which the results are averaged. By default,
no averaging is performed.
Note that invalid measurements (for example due to clipping or a missing fundamental
frequency) are not considered for averaging.
To clear the results for previous acquisitions used for averaging, select
.
Remote command:
6.3.5
Export Harmonic Results to File
Access: [MODE] = "Harmonic" > [FILE] > "Harmonics"
The results of a harmonic measurement can be exported similarly to a waveform (see
also
Chapter 12, "Documenting Results"
on page 265). The OneTouch functionality is
also available (see
Chapter 12.6, "Quick Save with OneTouch"
For details on the export file format, see
Chapter 6.3.6.2, "Export File Format for Har-
Harmonics Measurement (Option R&S
RTH-K34)