Remote Control
R&S
®
RTM20xx
339
User Manual 1317.4726.02 ─ 01
Suffix:
<m>
.
1..4
Selects the measurement place.
Parameters:
<SignalSource>
NONE | CH1 | CH2 | CH3 | CH4 | MA1 | MA2 | MA3 | MA4 | MA5 |
RE1 | RE2 | RE3 | RE4 TRIGger | CH1 | CH2 | CH3 | CH4 | MA1 |
MA2 | MA3 | MA4 | MA5 | RE1 | RE2 | RE3 | RE4 | TRIGger |
D0..D15 | D70 | D158
Waveform to be measured, required for all measurement types
CH1 | CH2 | CH3 | CH4
Active signal channels 1 to 4
MA1 | MA2 | MA3 | MA4 | MA5
Active math channels 1 to 5
RE1 | RE2 | RE3 | RE4
Active reference channels 1 to 4
TRIGger
Only return value. TRIG is returned if the measurement type is a
trigger measurement: TFRequency | TPERiode measure the A-
trigger source, TBFRequency | TBPeriod measure the B-trigger
source.
D0..D15
Active digital channels, if MSO option R&S
RTM-B1 is installed.
The following automatic measurements are possible: frequency,
period, edge and pulse counts, phase, delay, duty cycle, burst
width.
D70 | D158
Active digital pods D0...D7 and D8...D15, available if MSO option
R&S
RTM-B1 is installed.
*RST:
CH1
<ReferenceSource> NONE | CH1 | CH2 | CH3 | CH4 | MA1 | MA2 | MA3 | MA4 | MA5 |
RE1 | RE2 | RE3 | RE4 | NONE | CH1 | CH2 | CH3 | CH4 | MA1 |
MA2 | MA3 | MA4 | MA5 | RE1 | RE2 | RE3 | RE4 | D0..D15 | D70 |
D158
Second waveform, reference source that is required for delay and
phase mesurements.
MEASurement<m>:DELay:SLOPe
<SignalSlope>,<ReferenceSlope>
Sets the edges to be used for delay measurement. The associated waveforms are
defined with
Parameters:
<SignalSlope>
POSitive | NEGative
Slope of source 1 ( first waveform)
*RST:
POS
Command Reference