Remote Control
R&S
®
RTM20xx
292
User Manual 1317.4726.02 ─ 01
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Skew>
Deskew value
Default unit: s
CHANnel<m>:THReshold
<Threshold>
Threshold value for digitization of analog signals. If the signal value is higher than the
threshold, the signal state is high (1 or true for the boolean logic). Otherwise, the signal
state is considered low (0 or false) if the signal value is below the threshold.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Threshold>
Default values are:
TTL: 1.4 V
ECL: -1.3 V
CMOS: 2.5 V
*RST:
1.4
Default unit: V
CHANnel<m>:LABel
<Label>
Specifies a name for the selected channel.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Label>
String value
String with max. 8 characters, only ASCII characters can be used
CHANnel<m>:LABel:STATe
Shows or hides the channel name.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<State>
ON | OFF
*RST:
OFF
Command Reference