Remote Control
R&S
®
RTM20xx
291
User Manual 1317.4726.02 ─ 01
Parameters:
<BandwidthLimit>
FULL | B400 | B200 | B20
FULL
Use full bandwidth.
B400 | B200 | B20
Limit to 400MHz, 200 MHz, or 20 MHz respectively.
B400 is not available for 350 MHz instruments (R&S
RTM203x).
*RST:
FULL
CHANnel<m>:POLarity
<Polarity>
Turns the inversion of the signal amplitude on or off. To invert means to reflect the voltage
values of all signal components against the ground level. Inversion affects only the display
of the signal but not the trigger.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Polarity>
NORMal | INVerted
*RST:
NORM
CHANnel<m>:OVERload
<Overload>
Retrieves the overload status of the specified channel from the status bit. When the
overload problem is solved, the command resets the status bit.
Suffix:
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Overload>
ON | OFF
Use OFF to reset the overload status bit.
*RST:
OFF
Example:
CHANnel2:OVERload?
Queries the overload status of channel 2.
CHANnel2:OVERload OFF
Resets the overload status bit.
CHANnel<m>:SKEW
<Skew>
Skew or deskew compensates delay differences between channels caused by the dif-
ferent length of cables, probes, and other sources. Correct deskew values are important
for accurate triggering.
Command Reference