Putting into Operation
R&S
®
RT-ZD40
17
User Manual 1410.5363.02 ─ 05
4. Adjust the zero position of the waveform using the appropriate function of the
oscilloscope ("AutoZero", "Zero Adjust" or similar).
The waveform is set to 0 V on the horizontal centerline of the oscilloscope.
2.3.2
Micro Button
The micro button provides easy and quick access to important functions of the
Rohde
&
Schwarz oscilloscope. After a function has been assigned, pressing the
micro button remotely controls this specific function on the base unit. For exam-
ple, "Run continuous" or "Run single" are often assigned to the micro button.
The configuration of the micro button is part of the probe settings of the channel
to which the probe is connected. For more details, see the oscilloscope's user
manual.
2.4
Offset Compensation
The R&S
RT-ZD40 features two offset compensation functions: a differential off-
set compensation and a common mode offset compensation. Each function com-
pensates a particular DC component of the input signal, even in front of the active
amplifier in the probe tip. For a definition of differential and common mode input
voltages, see
Chapter 4, "Characteristics of Differential Probes"
2.4.1
Differential Offset
The differential offset compensation is often referred to as
offset compensation
.
Offset Compensation