Remote Control Commands
R&S
®
RTM3000
471
User Manual 1335.9090.02 ─ 08
TRIGger:EVENt:SCRSave
<SaveScreenshot>
If ON, the instrument saves a screenshot on each trigger event. The screenshot is
saved when the acquisition is complete. During saving, the acquisition stops, and
restarts when saving is finished. Thus, the waveform update rate decreases signifi-
cantly.
To set up the target directory, file name, color and file format, use the following com-
mands:
●
●
●
●
Parameters:
<SaveScreenshot>
ON | OFF
*RST:
OFF
TRIGger:EVENt:TRIGgerout
<PulseOut>
If ON, the instrument generates a pulse on the Aux
Out connector on trigger event.
The acquisition is not delayed, the pulse generation runs asynchronously.
Enabling "Pulse" sets the "Setup" > "Aux Out" selection to "Trigger Out", and vice
versa. Selecting another item in the "Aux Out" menu disables the pulse action.
To set the pulse width and polarity of the trigger out pulse, use:
●
●
Parameters:
<PulseOut>
ON | OFF
*RST:
OFF
TRIGger:EVENt:WFMSave
<SaveWaveform>
If ON, the instrument saves the data of a selected waveform on each trigger event. The
waveform is saved when the acquisition is complete. During saving, the acquisition
stops, and restarts when saving is finished. Thus, the waveform update rate decreases
significantly.
If fast segmentation is active, only the last acquisition is saved.
.
Set up the target directory, file name and file format, use
.
Parameters:
<SaveWaveform>
ON | OFF
*RST:
OFF
Trigger
www.allice.de