Remote control commands
R&S
®
RTB2000
559
User Manual 1333.1611.02 ─ 11
Return values:
<Condition>
Condition bits in decimal representation. ALIGnment (bit 0) ,
SELFtest (bit 1) , AUToset (bit 2), WTRigger (bit 3).
Range:
1 to 65535
Increment: 1
Usage:
Query only
STATus:OPERation:ENABle
<Enable>
Controls the ENABle part of the STATus:OPERation register. The ENABle defines
which events in the EVENt part of the status register are forwarded to the OPERation
summary bit (bit 7) of the status byte. The status byte can be used to create a service
request.
Parameters:
<Enable>
Range:
1 to 65535
Increment: 1
Example:
STATus:OPERation:ENABle 5
The ALIGnment event (bit 0) and AUToset event (bit 2) are for-
warded to the OPERation summary bit of the status byte.
STATus:OPERation:NTRansition
<NegativeTransition>
Parameters:
<NegativeTransition> Range:
1 to 65535
Increment: 1
STATus:OPERation:PTRansition
<PositiveTransition>
Parameters:
<PositiveTransition> Range:
1 to 65535
Increment: 1
STATus:OPERation[:EVENt]?
Return values:
<Event>
Range:
1 to 65535
Increment: 1
Usage:
Query only
15.14.2
STATus:QUEStionable registers
The commands of the
STATus:QUEStionable
subsystem control the status reporting
structures of the STATus:QUEStionable registers:
Status reporting