Remote control commands
R&S
®
RTB2000
519
User Manual 1333.1611.02 ─ 11
is set to
ERRCondition
.
Parameters:
<IdParityError>
ON | OFF
*RST:
OFF
Manual operation:
See
TRIGger:A:LIN:SYERror
<SyncError>
Triggers if synchronization caused an error.
is set to
ERRCondition
.
Parameters:
<SyncError>
ON | OFF
*RST:
OFF
Manual operation:
See
TRIGger:A:LIN:ICONdition
<IdentifierCondition>
Sets the comparison condition for the identifier: If the pattern contains at least one X
(don't care), you can trigger on values equal or not equal to the specified value. If the
pattern contains only 0 and 1, you can also trigger on a range greater than or lower
than the specified value.
is set to
ID
or
IDDT
.
Parameters:
<IdentifierCondition> EQUal | NEQual | GTHan | LTHan
*RST:
EQ
Manual operation:
See
TRIGger:A:LIN:IDENtifier
<Identifier>
Defines the identifier pattern.
is set to
ID
or
IDDT
.
Parameters:
<Identifier>
String containing binary pattern. Characters 0, 1, and X are
allowed. Enter the 6 bit identifier without parity bits, not the pro-
tected identifier.
Manual operation:
See
TRIGger:A:LIN:DATA
<Data>
Defines the data pattern. The number of bytes in the data pattern is defined with
.
Serial bus analysis