Remote control commands
R&S
®
RTB2000
506
User Manual 1333.1611.02 ─ 11
.................................................................................510
BUS<b>:CAN:FRAMe<n>:BYTE<o>:STATe?
.................................................................... 511
BUS<b>:CAN:FRAMe<n>:BYTE<o>:VALue?
....................................................................511
BUS<b>:CAN:FCOunt?
Returns the number of received frames.
Suffix:
<b>
.
1 | 2
Return values:
<FrameCount>
Total number of decoded frames.
Usage:
Query only
BUS<b>:CAN:FRAMe<n>:TYPE?
Returns the type of the specified frame.
Suffix:
<b>
.
1 | 2
<n>
*
Selects the frame (1...n).
Return values:
<FrameType>
DATA | REMote | ERR | OVLD
Data, remote, error or overload frame
Usage:
Query only
BUS<b>:CAN:FRAMe<n>:STATus?
Returns the overall state of the specified frame.
Suffix:
<b>
.
1 | 2
<n>
*
Selects the frame (1...n).
Return values:
<FrameStatus>
OK | BTST | CRCD | ACKD | CRC | EOFD | NOACk |
INSufficient
OK: frame is valid.
BTST: bit stuffing error occured
CRCD: wrong CRC delimiter occured
ACKD: Wrong ACK delimiter occured
CRC: cyclic redundancy check failed
EOFD: wrong end of frame
NOACk: acknowlegde is missing
Serial bus analysis