Remote control commands
R&S
®
RTB2000
496
User Manual 1333.1611.02 ─ 11
Return values:
<FrameCount>
Total number of decoded frames.
Usage:
Query only
BUS<b>:UART:FRAMe<n>:STARt?
BUS<b>:UART:RX:FRAMe<n>:STARt?
BUS<b>:UART:TX:FRAMe<n>:STARt?
Return the start time of the specified frame.
Suffix:
<b>
.
1 | 2
<n>
*
Selects the frame.
Return values:
<TxFrameStart>
Time in s, range depends on sample rate, record length, and
time base
Usage:
Query only
BUS<b>:UART:FRAMe<n>:STOP?
BUS<b>:UART:RX:FRAMe<n>:STOP?
BUS<b>:UART:TX:FRAMe<n>:STOP?
Return the end time of the specified frame.
Suffix:
<b>
.
1 | 2
<n>
*
Selects the frame.
Return values:
<TxFrameStop>
Time in s, range depends on sample rate, record length, and
time base
Usage:
Query only
BUS<b>:UART:FRAMe<n>:STATe?
BUS<b>:UART:RX:FRAMe<n>:STATe?
BUS<b>:UART:TX:FRAMe<n>:STATe?
Returns the status of the specified frame.
Suffix:
<b>
.
1 | 2
<n>
*
Selects the frame.
Return values:
<TxFrameState>
STER | SPER | PRER | BRE | OK | INS
Serial bus analysis