Remote control commands
R&S
®
RTB2000
417
User Manual 1333.1611.02 ─ 11
Set
OFF
to disable the measurement.
Suffix:
<m>
.
1..4
Sets the number of the meter measurement.
Parameters:
<MeasurementType> DC | ACDCrms | ACRMs | OFF
DC: mean value of the signal
ACDCrms: RMS value of the signal (AC+DC RMS)
ACRMs: RMS value of the signal's AC component (AC RMS)
OFF: disables the selected meter measurement.
*RST:
OFF
Manual operation:
See
DVM<m>:RESult[:ACTual]?
Returns the current value of the selected meter measurement.
Suffix:
<m>
.
1..4
Sets the number of the meter measurement.
Example:
DVM2:SOUR CH2
DVM2:TYPE DCRMs
DVM2:RES?
<-- 7.089E-01
An RMS measurement is performed on measurement place 2,
on channel 2. The result is 708.9
mV.
Usage:
Query only
DVM<m>:RESult[:ACTual]:STATus?
Returns the result value and the status of the result.
The status is the decimal representation of a 4-bit register value:
●
Bit 0 = 1: result is valid
●
Bit 1 = 1: no result available
●
Bit 2 = 1: clipping occurs
●
Bit 3 = 1: no period found
Suffix:
<m>
.
1..4
Sets the number of the meter measurement.
Return values:
<CurrentValue>
Measured value
<Status>
Decimal status value
Applications