Remote control commands
R&S
®
RTB2000
389
User Manual 1333.1611.02 ─ 11
MEASurement<m>:RESult:NPEak?
<NegativePeak>
Returns the minimum measurement result of the current measurement series.
Suffix:
<m>
.
1..6
Selects the measurement place.
Query parameters:
<NegativePeak>
Minimum measurement result
Usage:
Query only
MEASurement<m>:RESult:PPEak?
<PositivePeak>
Returns the maximum measurement result of the current measurement series.
Suffix:
<m>
.
1..6
Selects the measurement place.
Query parameters:
<PositivePeak>
Maximum measurement result
Usage:
Query only
MEASurement<m>:RESult:WFMCount?
<WaveformCount>
Returns the current number of measured waveforms.
The query for MIN and MAX values is not possible.
Suffix:
<m>
.
1..6
Selects the measurement place.
Query parameters:
<WaveformCount>
Number of measured waveforms
Usage:
Query only
MEASurement<m>:STATistics:WEIGht?
<BufferSize>
Returns the size of the statistics buffer.
Suffix:
<m>
.
1..6
Selects the measurement place.
Return values:
<BufferSize>
The buffer size is always 1000 result values.
Usage:
Query only
Measurements