Remote control commands
R&S
®
RTB2000
317
User Manual 1333.1611.02 ─ 11
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Offset>
Offset value
Range:
Depend on vertical scale and probe attenuation.
Increment: Depends on vertical scale and probe attenuation.
*RST:
0
Default unit: V
Manual operation:
See
"[Offset/Position (upper knob)]"
See
CHANnel<m>:COUPling
<Coupling>
Selects the connection of the indicated channel signal - coupling and termination.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Coupling>
DCLimit | ACLimit | GND
DCLimit
DC coupling passes the input signal unchanged.
ACLimit
Removes the DC offset voltage from the input signal.
GND
Connection to a virtual ground. All channel data is set to o V.
Manual operation:
See
See
CHANnel<m>:BANDwidth
<BandwidthLimit>
Selects the bandwidth limit for the indicated channel.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<BandwidthLimit>
FULL | B20
FULL
Use full bandwidth.
Waveform setup