Preparing for Use
R&S
®
OSP
‑
B157W8
(PLUS)
R&S
®
OSP
‑
B157WX
27
Getting Started 1178.5808.02 ─ 03
Risk of injury if feet are folded out
The feet can fold in if they are not folded out completely or if the instrument
is shifted, which can cause damage or injury.
●
Fold the feet completely in or out to ensure stability of the instrument.
Never shift the instrument when the feet are folded out.
●
When the feet are folded out, do not work under the instrument or place
anything underneath.
●
The feet can break if they are overloaded. The overall load on the fol-
ded-out feet must not exceed 100 N.
F
max
Risk of instrument damage due to insufficient airflow in a rack
If you mount several instruments in a rack, you need an efficient ventilation
concept to ensure that the instruments do not overheat. Insufficient airflow
for a longer period can disturb the operation and even cause damage.
For putting the modules into operation after they are mounted inside a switch unit
and optionally inside a rack, proceed as described in the
of the
R&S
OSP. This manual also contains valuable information on the switch unit's
setup, on remote operation, on working with the operating system, and much
more.
Putting into Operation