Instrument Functions
R&S
®
NGL200/NGM200
59
User Manual 1178.8736.02 ─ 09
Figure 6-14: Overcurrent protection dialog
2. Activate the "Enabled" menu item.
The R&S
NGL/NGM enables the OCP and displays the OCP icon on the selected
channel status bar information.
3. Set the required "Fuse Delay Time" and "Fuse Delay At Output-On".
The R&S
NGL/NGM displays the on-screen keypad to set the values.
● "Fuse Delay Time": The time taken to turn off the affected channel after OCP is
triggered.
● "Fuse Delay At Output-On": The time taken after channel output is turned on
before OCP is put into operation.
4. Confirm value with the unit key (ms or s).
5. Activate the "Link to Channel 2" or "Link to Channel 1" menu item.
● ON: The linked channel is turned off when an OCP event is triggered.
● OFF: The linked channel is not affected when an OCP event is triggered.
6.5.2
Overvoltage Protection (OVP)
When the output voltage exceeds the limit set for the respective channel, an alert is
triggered and the affected channel is turned off according to the settings configured in
the OVP dialog.
1. Select "Overvoltage Protection (OVP)" from the menu.
The R&S
NGL/NGM displays the OVP dialog.
Protection