Instrument functions
R&S
®
LCX Series
109
User Manual 1179.2260.02 ─ 02
Example: Binning configuration data for a resistor measurement
The example defines a binning measurement that returns the measurement results for
the resistance and impedance parameter pair (
Rs-X
), specified for 4 measurement cat-
egories. The R&S
LCX returns the nominal value as absolute value, and the deviation
in percent (
NOMPCT)
.
#Signal,BOTH
#Measurement,Rs-X
#PrimaryFormat,NOMPCT
#SecondaryFormat,ABS
Pri0,Pri1
1e3,0.1
1e3,1
1e3,5
1e3,10
Loading the binning file
To upload the binning, proceed as follows:
1. Connect the USB stick to the USB B connector at the front panel of the instrument.
2. To access the binning settings dialog, select [settings] > "Functions" > "Binning".
3. Select "Load from File".
4. Select the partition "USB1A".
If a binning folder exists on the memory stick, the R&S
LCX leads you to this direc-
tory automatically.
5. Select the corresponding
*.csv
binning file.
The R&S
LCX indicates the file properties of the selected file on the right.
6. Confirm your selection with "Load".
Returned to the "Binning" dialog, the instrument indicates the selected file and
memory location.
Specific instrument functions
www.allice.de
Allice Messtechnik GmbH