R&S FSP
Maintenance and Instrument Interfaces
Instrument Interfaces
8.8
Operating Manual 1164.4556.12 - 06
•
GND
Interface ground, connected to instrument ground
•
DSR
(Data Set Ready)
Input, active LOW,
Indicates that the remote station is ready to receive data.
•
RTS
(Request To Send)
Output, active LOW.
Indicates that the local terminal wants to transmit data.
•
CTS
(Clear To Send)
Input, active LOW.
Used to tell the local terminal that the remote station is ready to receive data.
•
RI
(Ring Indicator)
Not used in
R&S FSP.
Input, active LOW.
Used by a modem to indicate that a remote station wants to establish a
connection.
8.3.3.3
Transmission Parameters
To ensure error-free data transmission, the parameters of the instrument and the
controller must have the same settings. The parameters are defined in the SETUP-
GENERAL SETUP menu.
Example
Transmission of character 'A' (41 hex) in 7-bit ASCII code, with even parity and 2
stop bits:
Transmission rate
(baud rate)
The following transmission rates can be set in the
instrument: 110, 300, 600, 1200, 2400, 4800, 9600, 19200,
38400, 57600, 115200, 138000.
Data bits
Data transmission is in 7- or 8-bit ASCII code. The first bit
transmitted is the LSB (least significant bit).
Start bit
Each data byte starts with a start bit. The falling edge of the
start bit indicates the beginning of the data byte.
Parity bit
In order to detect errors, a parity bit may be transmitted. No
parity, even parity or odd parity may be selected. In addition,
the parity bit can be set to logic '0' or to logic '1'.
Stop bits
The transmission of a data byte is terminated by 1, 1,5 or 2
stop bits.
03
04
05
06
07
08
09
10
11
02
01
Bit 01
Start bit
Bit 02...08
Data bits
Bit 09
Parity bit
Bit 10...11
Stop bits
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...