Operating Manual 1164.4556.12 - 06
6.113
R&S FSP
Remote Control – Description of Commands
DISPlay Subsystem
DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:SCALe]:RLEVel:OFFSet
-
200dB to 200dB
This command defines the offset of the reference level in the selected measurement window.
Depending on the coupling of the measurement windows, it is valid for both screens or only for the
selected measurement window (
).
The numeric suffix at TRACe<1...3> is irrelevant.
Example:
"DISP:TRAC:Y:RLEV:OFFS -10dB"
Characteristics:
*RST value: 0dB
SCPI: conform
Mode:
A
DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:SCALe]:RPOSition
0 to 100PCT
This command defines the position of the reference value in the selected measurement window. The
numeric suffix in TRACe<1...3> is irrelevant.
In NETWORK mode (Tracking Generator/Ext. Generator Option R&S FSP-B9/R&S FSP-B10) with
active normalization, RPOSition defines the reference point for the output of the normalized
measurement results.
Example:
"DISP:TRAC:Y:RPOS 50PCT"
Characteristics:
*RST value:
100PCT(analyzer mode)
50 PCT (NETWORK mode)
SCPI: conform
Mode:
A
DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:SCALe]:RVALue
<numeric_value>
If the tracking generator option or the external generator control option (R&S FSP-B9/R&S FSP-B10)
is available and the normalization in the NETWORK mode is activated, this value defines the power
value assigned to the reference position in the selected measurement window. This value corresponds
to the parameter REFERENCE VALUE in manual operation.
The numeric suffix at TRACe<1...3> is irrelevant.
Example:
"DISP:TRAC:Y:RVAL 0"
'Sets the power value assigned to the reference position to 0 dB (Tracking
Generator/Ext. Generator Control option).
Characteristics:
*RST value:
0 dB (NETWORK mode)
SCPI: device-specific
Mode:
A
Содержание R&S FSP Series
Страница 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Страница 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Страница 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Страница 362: ...R S FSP Instrument Functions Trigger Port Option R S FSP B28 4 290 Operating Manual 1164 4556 12 06 ...
Страница 764: ...R S FSP Remote Control Programming Examples More Complex Programming Examples 7 30 Operating Manual 1164 4556 12 06 ...
Страница 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Страница 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Страница 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Страница 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...