Remote Commands
R&S
®
Spectrum Rider FPH
308
User Manual 1321.1011.02 ─ 06.00
Parameters:
<Format>
ASCii | REAL
ASCii
Returns the data in ASCII format, separated by commas.
REAL
Returns the data as 32-bit IEEE 754 floating point numbers in
the "definite length block format".
*RST:
ASCii
In REAL, 32 format, a string of return values would look like:
#42424<value 1><value 2>...<value n>
with
#4 - representing the number of digits of the following number of
data bytes (= 4 in this example);
2524 - representing the number of following data bytes (2524,
corresponds to the 711 sweep points of the R&S Spectrum
Rider.
<value> - representing 4-byte floating point value.
Example:
FORM ASC
Selects the ASCii data format.
[SENSe:]DETector<1...2>[:FUNCtion]
<Detector>
This command selects a detector function.
The numeric suffix at DETector specifies the primary or secondary detector.
Suffix:
<1...2>
.
1...2
Parameters:
<Detector>
POSitive | NEGative | SAMPle | RMS | AVERage | APEak
*RST:
APE
For more information, see chapter "Detectors".
Example:
DET POS
Sets the primary detector to "positive peak".
[SENSe:]DETector<1...2>[:FUNCtion]:AUTO
<State>
This command couples and decouples the detector to the trace mode.
The numeric suffix at DETector specifies the primary or secondary detector.
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
DET:AUTO OFF
Turns off automatic detector selection for the primary detector.
Remote Control - Commands