NB-IoT NPRACH Measurement
R&S
®
CMW-KM300/-KS300
378
User Manual 1178.5689.02 ─ 06
READ:NIOT:MEAS<i>:PRACh:MODulation:CURRent?
READ:NIOT:MEAS<i>:PRACh:MODulation:AVERage?
READ:NIOT:MEAS<i>:PRACh:MODulation:SDEViation?
Return the current, average and standard deviation single value results.
The values described below are returned by
FETCh
and
READ
commands.
CALCulate
commands return limit check results instead, one value for each result listed below.
Return values:
<1_Reliability>
<2_OutOfTol>
Out of tolerance result, i.e. percentage of measurement inter-
vals of the statistic count for modulation measurements exceed-
ing the specified modulation limits.
Default unit: %
<3_EVM_RMSlow>
EVM RMS value, low EVM window position
Default unit: %
<4_EVM_RMShigh> EVM RMS value, high EVM window position
Default unit: %
<5_EVMpeakLow>
EVM peak value, low EVM window position
Default unit: %
<6_EVMpeakHigh>
EVM peak value, high EVM window position
Default unit: %
<7_MErr_RMSlow>
Magnitude error RMS value, low EVM window position
Default unit: %
<8_MErr_RMShigh> Magnitude error RMS value, high EVM window position
Default unit: %
<9_MErrPeakLow>
Magnitude error peak value, low EVM window position
Default unit: %
<10_MErrPeakHigh> Magnitude error peak value, high EVM window position
Default unit: %
<11_PErr_RMSlow> Phase error RMS value, low EVM window position
Default unit: deg
<12_PErr_RMSh>
Phase error RMS value, high EVM window position
Default unit: deg
<13_PErrPeakLow> Phase error peak value, low EVM window position
Default unit: deg
<14_PErrPeakHigh> Phase error peak value, high EVM window position
Default unit: deg
<15_FreqError>
Carrier frequency error
Default unit: Hz