NB-IoT Signaling
R&S
®
CMW-KM300/-KS300
144
User Manual 1178.5689.02 ─ 06
Firmware/Software:
V3.7.20
Manual operation:
See
SENSe:NIOT:SIGN<i>:UL:ADPower:PATHloss?
Queries the pathloss resulting from the advanced UL power settings.
Return values:
<Pathloss>
Default unit: dB
Example:
See
Usage:
Query only
Firmware/Software:
V3.7.20
Manual operation:
See
CONFigure:NIOT:SIGN<i>:UL:ADPower:PIRPower
<TargetPower>
Specifies the "preambleInitialReceivedTargetPower" value, signaled to the UE if
advanced UL power configuration applies.
Parameters:
<TargetPower>
Range:
-120 dBm to -90 dBm
Increment: 2 dB
*RST:
-104 dBm
Default unit: dBm
Example:
See
Firmware/Software:
V3.7.20
Manual operation:
See
"Preamble Initial Received Target Power"
CONFigure:NIOT:SIGN<i>:UL:ADPower:PNNPusch
<P0NomNPUSCH>
Specifies the "p0-NominalNPUSCH" value, signaled to the UE if advanced UL power
configuration applies.
Parameters:
<P0NomNPUSCH>
Range:
-126 dBm to 24 dBm
*RST:
-115 dBm
Default unit: dBm
Example:
See
Firmware/Software:
V3.7.20
Manual operation:
See
Command Reference