NB-IoT Signaling
R&S
®
CMW-KM300/-KS300
112
User Manual 1178.5689.02 ─ 06
Setting parameters:
<Control>
OFF | ON
Switch generator
ON
or
OFF
*RST:
OFF
Return values:
<GeneratorState>
OFF | PENDing | ON
OFF:
generator switched off
PEND:
generator switched on but no signal available yet
ON:
generator switched on, signal available
*RST:
OFF
Example:
See
Firmware/Software:
V3.7.20
Manual operation:
See
SOURce:NIOT:SIGN<i>:CELL:STATe:ALL?
Returns detailed information about the "NB-IoT Signaling" generator state.
Return values:
<MainState>
OFF | ON
OFF:
generator switched off
ON:
generator switched on
<SyncState>
PENDing | ADJusted
PENDing:
generator turned on (off) but signal not yet (still) avail-
able
ADJusted:
physical output signal corresponds to main genera-
tor state
Example:
See
Usage:
Query only
Firmware/Software:
V3.7.20
Manual operation:
See
"ON | OFF (key) / NB-IoT Signaling (softkey)"
SENSe:NIOT:SIGN<i>:RRCState?
Queries whether an RRC connection is established (connected) or not (idle).
Return values:
<State>
IDLE | CONNected
Example:
See
Usage:
Query only
Firmware/Software:
V3.7.20
Manual operation:
See