Modules
R&S
®
OSP
146
User Manual 1178.7100.02 ─ 05
3.59
R&S
OSP-PM-I Power Sensor Integration
Figure 3-143: R&S
OSP-PM-I
R&S
OSP-PM-I (order no. 1515.5985.02) allows integrating a power sen-
sor into an R&S
OSP switch unit. The module has an N-type feedthrough connector
and a USB connector, both mounted in the module's front panel.
Note:
150 N
·
cm
for
N
connectors.
The R&S
OSP-PM-I requires a separately ordered power sensor of the
Rohde
&
Schwarz product family
. Without ordering a sensor, it is not
included in the module delivery. We recommend using a power sensor version with a
short cable, for example the
(order no. 1417.0409.04).
The module R&S
OSP-PM-I is designed with a base plate that carries and fixes the
power sensor and its cable. The sensor's RF input N connector is directly screwed to
the feedthrough, while the sensor's control and output connector (Odumac L-series
with 6 pins) is connected to a USB adapter. This adapter includes a lowpass filter and
provides a USB standard B connector in the module's front panel.
The power sensor is entirely controlled via the USB interface, it has no connection to
the switch unit's motherboard or internal connectors. Hence, this module does not
require a module bus cable. It only serves for housing the power sensor inside the
switch unit.
Figure 3-144: Pinout and block diagram of the power sensor integration module R&S
OSP-PM-I
For remote control of this module, refer to the
of your power sensor.
R&S
OSP-PM-I Power Sensor Integration