SINC
Returns the trace data as the result of a SINC interpolation
of the acquired samples. This setting is only reasonable if
the selected number of trace points is higher than the num-
ber of samples which results from the internal sample rate
in the selected trace time. The SINC interpolation recon-
structs the original signal sequence best from the sampled
values. Use it to get a smooth resulting curve.
*RST:
SINC
Example:
SENS:TRAC:TIME 100E-6
Sets the trace time to 100 µs. With an internal sample rate
of 2 MHz, the sensor module performs 200 measurements.
SENS:TRAC:POIN 400
SENS1:TRACe:UPS:TYPE SINC
If you select the number of trace points higher than the
number of physical measurements, you achieve a smooth
result curve using the
SINC
interpolation.
SENS:TRAC:POIN 200
SENS1:TRACe:UPS:TYPE SINC
If the trace points are up to the number of physical mea-
surements, the
SINC
interpolation does not affect the trace
results.