Instrument functions
R&S
®
NGP800
95
User Manual 5601.5610.02 ─ 09
Figure 6-63: Update device dialog
2. Select the source and file location to update instrument firmware.
3. Select "UPDATE" to update the instrument firmware.
The R&S
NGP800 updates the instrument firmware accordingly.
6.17
Adjustment
Adjustment shall be done at ambient temperature of 25
°
C
±
2
°
C .
The instrument must be operated for at least 30 minutes before executing the adjust-
ment .
Thick wires are recommended for connecting the shunt resistor to avoid huge voltage
drop and excessive heating.
For ease of maintenance, the R&S
NGP800 provides two adjustment procedures
which you can apply on the instrument:
●
Chapter 6.17.1, "Analog In adjustment"
●
Chapter 6.17.2, "Channel adjustment"
► Press [Settings] key.
The R&S
NGP800 displays the device/channel menu window.
Adjustment