Remote control commands
R&S
®
NGP800
170
User Manual 5601.5610.02 ─ 09
Usage:
Query only
HCOPy:SIZE:Y?
Returns the vertical dimension of the screenshots.
Usage:
Query only
LOG[:STATe]
<arg0>
Sets or queries the data logging state.
Parameters:
<state>
1
Data logging function is enabled.
0
Data logging function is disabled.
*RST:
0
Example:
LOG ON
LOG?
-> 1
Data logging function is activated.
LOG:COUNt
<Set new value>[, <Return min or max>]
LOG:COUNt?
[<Return min or max>]
Sets or queries the number of measurement values to be captured.
Setting parameters:
<count>
<numeric value> | MIN | MINimum | MAX | MAXimum
<numeric value>
Number of measurement values to be captured is set in the
range of 1 to 10000000.
MIN | MINimum
Minimum number of measurement values to be captured is set
at 1.
MAX | MAXimum
Maximum number of measurement values to be captured is set
at 10000000.
Parameters for setting and query:
<count>
MIN | MINimum | MAX | MAXimum
Returns the number of measurement values.
Example:
LOG:COUN MAX
LOG:COUN? MAX
-> 10000000
Data and file management commands