VOLTage:OFFSet? [MINimum | MAXimum]
Queries the output offset value.
Return values
:
Numeric value in V (Volt) depending on the amplitude setting.
MIN
:
- 1.01
MAX
: 1.01
DATA {<binary block>}
Defines the arbitrary waveform. The instrument can handle data as binary values. In the binary
block format, a block header precedes the waveform data. The block header has the following
format:
The instrument represents binary data as 16-bit integers, which are sent as two bytes. Therefore,
the total number of bytes is always twice the number of data points in the waveform and must
always be an even number. For example, 512 bytes are required to download a waveform with
256 points.
DATA:MODulation {<Binary Block>}
Defines and transfers an arbitrary modulation waveform. The instrument can handle data as
binary values. Please refer to the DATA {<binary block>} command.
DATA:SAVE „Arbitrary Name“
Transfers the previously generated arbitrary waveform from RAM (random access memory) into
the non-volatile memory (ROM). The generated file is accessable via the internal file system
afterwards.
Example
:
DATA:SAVE „ABC.HRT“
NOTICE
You are only able to save an USB stick file in HRT format into the internal memory of
the HMF intrument.
#
3
512
Start of
data block
Number of digits
to follow
Even number of bytes to follow
[512 bytes = 256 points]