Measurements
R&S
®
FSW
164
User Manual 1173.9411.02 ─ 19
Power Mode
The measured power values can be displayed directly for each trace ("Clear/Write"), or
only the maximum values over a series of measurements can be displayed ("Max
Hold"). In the latter case, the power values are calculated from the current trace and
compared with the previous power value using a maximum algorithm. The higher value
is retained. If "Max Hold" mode is activated, "Pwr Max" is indicated in the table header.
Note that the
trace
mode remains unaffected by this setting.
Remote command:
Optimized Settings (Adjust Settings)
All instrument settings for the selected channel setup (channel bandwidth, channel
spacing) can be optimized automatically.
The adjustment is carried out only once. If necessary, the instrument settings can be
changed later.
The following settings are optimized by "Adjust Settings":
●
●
●
●
●
Note:
The reference level is not affected by this function. To adjust the reference level
Setting the Reference Level Automatically (Auto Level)
in the AUTO SET menu.
Remote command:
5.3.5.2
MSR Sub block and Tx Channel Definition
The "Tx Channel Settings" tab in the "MSR ACLR Setup" dialog box provides all the
channel settings to configure sub blocks and Tx channels in MSR ACLR measure-
ments.
Channel Power and Adjacent-Channel Power (ACLR) Measurement