Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
93
User Manual 1173.9286.02 ─ 15
Parameters:
<Frequency>
Range:
0 to fmax
*RST:
fmax/2
Default unit: Hz
f
max
is specified in the data sheet. min span is 10 Hz
Example:
FREQ:CENT 100 MHz
Defines a nominal frequency of 100 MHz.
Manual operation:
See
[SENSe:]POWer:RLEVel
<Power>
This command defines the nominal level.
Parameters:
<Power>
Numeric value in dBm.
Range:
-200 to 200
*RST:
0
Example:
POW:RLEV -20
Defines a nominal level of -20 dBm.
Manual operation:
See
INPut:ATTenuation
<Attenuation>
This command defines the total attenuation for RF input.
If you set the attenuation manually, it is no longer coupled to the reference level, but
the reference level is coupled to the attenuation. Thus, if the current reference level is
not compatible with an attenuation that has been set manually, the command also
adjusts the reference level.
Parameters:
<Attenuation>
Range:
see data sheet
Increment: 5 dB
*RST:
10 dB (AUTO is set to ON)
Example:
INP:ATT 30dB
Defines a 30 dB attenuation and decouples the attenuation from
the reference level.
Manual operation:
See
"Mechanical Attenuator / Value"
INPut:ATTenuation:AUTO
<State>
This command couples or decouples the attenuation to the reference level. Thus, when
the reference level is changed, the R&S
FSW determines the signal level for optimal
internal data processing and sets the required attenuation accordingly.
Configuring the Frontend