Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
87
User Manual 1173.9286.02 ─ 15
LAYout:ADD[:WINDow]?
<WindowName>,<Direction>,<WindowType>
This command adds a window to the display in the active measurement channel.
This command is always used as a query so that you immediately obtain the name of
the new window as a result.
To replace an existing window, use the
Parameters:
<WindowName>
String containing the name of the existing window the new win-
dow is inserted next to.
By default, the name of a window is the same as its index. To
determine the name and index of all active windows, use the
<Direction>
LEFT | RIGHt | ABOVe | BELow
Direction the new window is added relative to the existing win-
dow.
<WindowType>
text value
Type of result display (evaluation method) you want to add.
See the table below for available parameter values.
Return values:
<NewWindowName> When adding a new window, the command returns its name (by
default the same as its number) as a result.
Example:
LAY:ADD? '1',LEFT,MTAB
Result:
'2'
Adds a new window named '2' with a marker table to the left of
window 1.
Usage:
Query only
Table 8-3: <WindowType> parameter values for Phase Noise application
Parameter value
Window type
FDRift
Frequency drift
MTABle
Marker table
PNOise
Phase noise diagram
RNOise
Residual noise table
SNOise
Spot noise table
SPECtrum
Spectrum monitor
SPURs
Spur list
SRESults
Sweep result list
STABility
Frequency and level stability indicator
Configuring the Result Display