Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
84
User Manual 1173.9286.02 ─ 15
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
[SENSe:]SWEep:COUNt
<SweepCount>
This command defines the number of sweeps that the application uses to average
traces.
In case of continuous sweep mode, the application calculates the moving average over
the average count.
In case of single sweep mode, the application stops the measurement and calculates
the average after the average count has been reached.
Suffix:
<n>
.
Parameters:
<SweepCount>
If you set a sweep count of 0 or 1, the R&S
FSW performs one
single sweep.
Range:
0 to 200000
*RST:
200
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation:
See
[SENSe:]SWEep:FHDecade
This command stops the measurement in the current half decade and continues mea-
suring in the subsequent half decade.
Example:
SWE:FHD
Aborts the current measurement and continues in the next half
decade.
Usage:
Event
Manual operation:
See
Performing Measurements