Analysis
R&S
®
FSW-K40
55
User Manual 1173.9286.02 ─ 15
Preset All Traces
Resets all traces to their default mode.
Trace 1 - 2 mode = Clear Write; Trace 3 - 6 mode = Blank
Trace 1 smoothing = On, Trace 2 - 6 smoothing = Off
Set Trace Mode
Max | Avg | Min
Trace 1 mode: Max Hold
Trace 2 mode: Average
Trace 3 mode: Min Hold
Set Trace Mode
Max | ClrWrite | Min
Trace 1 mode: Max Hold
Trace 2 mode: Clear Write
Trace 3 mode: Min Hold
Trace Export ← Trace Config
The "Trace Export" tab contains functionality to export trace data.
"Export all
traces and all
table results"
Turns the export of all measurement results (traces and numerical
results) on and off.
If on, selecting a particular trace to export in the "Trace to Export"
dropdown menu is unavailable.
"Include Instru-
ment Measure-
ment Settings"
Includes or excludes the measurement configuration as shown in the
channel bar from the export.
"Trace to
Export"
Selects the trace that will be exported to a file.
"Decimal Sep-
arator"
Selects the decimal separator for floating-point numerals for the
ASCII Trace export. Evaluation programs require different separators
in different languages.
"Export Trace
to ASCII File"
Opens a file selection dialog box and saves the selected trace in
ASCII format to the specified file and directory.
Remote command:
Decimal separator:
Export trace to ASCII file:
Selecting a trace:
Export the header:
Copy Trace ← Trace Config
Access
: "Overview" > "Analysis" > "Traces" > "Copy Trace"
Or: TRACE > "Copy Trace"
Copies trace data to another trace.
The first group of buttons (labeled "Trace 1" to "Trace 6" ) selects the source trace. The
second group of buttons (labeled "Copy to Trace 1" to "Copy to Tace 6" ) selects the
destination.
Configuring Graphical Result Displays