Configuration
R&S
®
FSW-K40
38
User Manual 1173.9286.02 ─ 15
●
input coupling
After initial setup, the parameters for the measurement channel are stored upon exiting
and restored upon re-entering the channel. Thus, you can switch between applications
quickly and easily.
Apart from these settings, the following default settings are activated directly after a
measurement channel has been set to the Phase Noise application, or after a channel
preset:
Table 5-1: Default settings for phase noise measurement channels
Parameter
Value
Attenuation
Auto (0 dB)
Verify frequency & level
On
Frequency & level tracking
Off
Measurement range
1 kHz ... 1 MHz
Sweep type
Normal
X axis scaling
Measurement range
Y axis scaling
20 dBc/Hz ... 120 dBc/Hz
Smoothing
1%
Smoothing type
Linear
5.3 Configuring the Frontend
Access
: "Overview" > "Input / Frontend"
The "Frontend" tab of the "Measurement Settings" dialog box contains all funtions nec-
essary to configure the frontend of the RF measurement hardware.
Functions to configure the RF input described elsewhere:
●
Chapter 5.7.1, "Input Source Configuration"
Configuring the Frontend