Measurements and Result Displays
R&S
®
FSW-K40
16
User Manual 1173.9286.02 ─ 15
Type
Shows the number of the trace that is integrated (T[x]).
Start / Stop Offset
Shows the start and stop offset of the integration interval.
Int. PHN
Shows the Integrated Phase Noise.
The integral is calculated over the frequency range defined by the Start and Stop
Offset values.
PM
Shows the Residual PM result in degrees and rad.
FM
Shows the Residual FM results in Hz.
Jitter
Shows the Jitter in seconds.
For more information on residual noise see
Chapter 4.2, "Residual Effects"
Remote command:
Querying Residual PM:
Querying Residual FM:
Querying Jitter:
Querying Integrated Phase Noise
Querying user ranges:
FETCh:PNOise<t>:USER<range>:RFM?
FETCh:PNOise<t>:USER<range>:RMS?
FETCh:PNOise<t>:USER<range>:RPM?
Spot Noise
Spot noise is the phase noise at a particular frequency offset (or spot) that is part of the
measurement range. It is thus like a fixed marker.
The unit of spot noise results is dBc/Hz. The application shows the results in a table.
The table consists of a variable number of 10
x
frequencies (depending on the mea-
surement range), and a maximum of five user frequencies, with each row containing
the spot noise information for a particular frequency offset.
The spot noise information is made up out of several variables.