Remote Control Commands for Phase Noise Measurements
R&S
®
FSW-K40
140
User Manual 1173.9286.02 ─ 15
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:MARK3 ON
Switches on marker 3.
Manual operation:
See
See
CALCulate<n>:MARKer<m>:TRACe
<Trace>
This command selects the trace the marker is positioned on.
Note that the corresponding trace must have a trace mode other than "Blank".
If necessary, the command activates the marker first.
Suffix:
<n>
.
<m>
Parameters:
<Trace>
1 to 6
Trace number the marker is assigned to.
Example:
CALC:MARK3:TRAC 2
Assigns marker 3 to trace 2.
Manual operation:
See
" Assigning the Marker to a Trace "
CALCulate<n>:MARKer<m>:X
<Position>
This command moves a marker to a particular coordinate on the x-axis.
If necessary, the command activates the marker.
If the marker has been used as a delta marker, the command turns it into a normal
marker.
Suffix:
<n>
.
<m>
Parameters:
<Position>
Numeric value that defines the marker position on the x-axis.
Default unit: Hz
Example:
CALC:MARK2:X 1.7MHz
Positions marker 2 to frequency 1.7 MHz.
Using Markers